Fully automatic AED CardiAid CT0207RF
cardiaid_straight_on_open
Defibrillation schedule
Operation Mode: Fully Automatic
Shockwave: Biphasic
Energy issue: is adapted to the impedance or resistance of the patient
Energy issue – adults mode: 170J (average 50 Ω at low energy level)
270J (average 50 Ω at a high energy level)
Energy issue – Child mode: 50J (average 50 Ω at low energy level)
75J (average 50 Ω at a high energy level)
Shock pattern: constant or constructive
Charging time: Typically <10 seconds * Maximum <15 seconds *
ECG Analysis
Analysis time: Typically <10 seconds *
Derivation: II
Asystole threshold: <0.160 mV
Specificities NSR / Asystole: > 95 **
Sensitivity VF / PVT: > 90 **
Reaction to implanted pacemaker: Normal cardiac pacemaker rhythm is not detected as shockable
Motion detection: Check signal quality, giving audible warning if patient moves.
Indicators
Operating units: AED begins automatically when opened
Info button: Info button shows the elapsed time and number of shocks given since the device was opened
Icons: Clear and intuitive icons display (through clear illuminated steps)
Device status indication: Flashing symbols shows the status of the unit
Defibrillation Electrodes
Possible applications: Adults electrodes for patients > 8 years/25kg. Child electrodes suitable for patients of 1-8 years/<25kg.
Supplied: Disposable, self-adhesive, single-use electrodes, ready to use, sealed and packaged with the connector outside the package
Polarization: Not polarized, exchange is possible
Active surface area: 166cm2 (adults), 86cm2 (children)
Length of the cable: 130 cm
Life if unused: 36 months after production date
Data Management
Documentation of an event: Automatic recording of ECG and effort data are stored up to a frequency of 4 times with a maximum recording time of 4 hours.
Data Transfer: Via Bluetooth
Event review and device programming: Via Bluetooth connection to a PC with CardiAid Monitor Software (only for authorized service partners)
HomeTest
Schedule: Automatic, daily, monthly and when the valve is opened
Timing: Programmable
Scope vd test: Battery, Electronics, Software and Loading of the capacitor
Main battery
Type: Alkaline
Capacity: Up to 210 shocks *
Ability to monitor: Up to 20 hours * hours
Standby time: 2 years
Environmental conditions
Operation: i) Temperature: +5°C to + 55˚C

ii) Humidity: 0-95%
iii) Pressure: 572-1060h / PA
Storage and transport: i) Temperature: +5°C to + 55˚C
ii) Temperature: -20˚C to + 70˚C (max 2 weeks, without batteries and electrodes).
ii) Humidity: 0-95%
iii) Pressure: 500 -1160h / PA
Standards
Classification: IIb (93/42 / EEC)
Resuscitation protocol: ERC, ILCOR 2010
Standards: EN 60601-2-4: 3rd edition, rare use,
EMC: DIN EN 60601-1-2: 2007-12
Free fall: IEC 601-1: 1988 + A1: 1991 + A2: 1995
Protection: IEC 529: IP55 (protection against dust and spray water proof)
Dimensions and weight
Dimensions: 301 x 304 x 112mm
Weight: 3.1 kg (including battery and electrodes)

* Projected data is based on 20 ˚C with new fully-charged batteries. Values may vary depending on storage and environmental conditions, frequency of use, preconfigured settings and the life of the device

** The algorithm and the board of the CT0207 AEDs deliver the optimum combination of sensitivity and specificity. (Published in “Automated analysis of electrical signals of the human body for detecting or life threatening cardiac abnormalities,” Igor Tchoudovski, Mensch and Buch Verlag, ISBN 3-89820-984-9, Pg. 141)
Cardia International A/S, in its sole discretion, reserves the right to make changes related to the increase in technical features and accessories of the appliance. Cardia International A/S can not be held liable under any circumstances for errors in spelling and/or writing at all informative data, documents, images such as data or damage caused by direct, indirect, contingency, secondary legal circumstances.